1、Validation of Washer-Disinfectors in a Danish Hospital Setting-Cleaning Efficacy and StandardizationPeter Rubak,Assistant ProfessorClinical Nursing Research UnitAalborg University Hospital,DenmarkWhere I come fromDistance:Denmark Hong Kong:8746 kmNew CSSDWhere to start the validation processWhat the
2、y doWhat they bringQuality Control Specialist Hands-on experienceProcess EngineerProcess flow overviewResearchersKnowledge of evidenceWhat to test before the validation process?ORMachinecleaningCSSDRinsingLoadingManuel cleaningUltrasonic cleaningClean areaWhat to test before the validation process?O
3、RCSSDClean areaRinsing+ultrasonic cleaningWhat to test before the validation process?ORCSSDClean areaLoadingVSWagnerAiconAesculap(JG222R)What to test before the validation process?ORCSSDClean areaMachine cleaningPre-rinseWashingInt-rinse 1Int-rinse 2Pre-rinseWashingInt-rinseMediClean-AdvancedProzyme
4、Alka-XStandardization-Choosing instrumentsFor example:Scissors Tweezer Forceps NeedleholdersReal-life instruments used at ORDiscarded instruments from basic traysStandardization Placing the instrumentsDCBAEDCAEWhat to test and how?AcceptableAlertNot acceptableQuality limitsg/Instrument 200Test seque
5、nceTwo baskets per shelfThree baskets per shelfAdvanced washing protocolTwo baskets per shelfThree baskets per shelfAlka-X washing protocolTwo baskets per shelfThree baskets per shelfAdvanced wash+SonicationTwo baskets per shelfThree baskets per shelfAlka-X wash+SonicationTest resultsWagner Aicon Ae
6、sculapInstrument 1-A2 06Instrument 1-C197 012Instrument 1-D85242262Instrument 1-E15829171Instrument 3-A 0 04Instrument 3-BInstrument 3-C7612895Instrument 3-D15722062Instrument 3-E53186222Instrument 5-A1083136Instrument 5-BInstrument 5-C38296Instrument 5-D8668Instrument 5-E434927OP 126OP 262OP 349Two